Key takeaways
- Choose an analyser that meets the standard BS 7967 guidance refers to, BS EN 50379-3, and suits the appliances you work on.
- Follow the manufacturer's recalibration interval, typically annual, and keep every certificate.
- Stagger calibration dates across the team so you are never short of analysers in winter.
- Do the start-up checks every time: fresh-air zero, filters, water trap and probe hose.
- Record readings against the job and the analyser serial number, so any result can be traced.
Why the analyser sits at the centre of gas work
Only engineers in a class of persons approved by the Health and Safety Executive, in practice the Gas Safe Register, may work on gas fittings[1]. Much of that work relies on one instrument. The flue gas analyser confirms that an appliance is burning correctly, checks for carbon monoxide in the flue products and, where needed, in the room. The NHS lists gas boilers, gas cookers and gas or paraffin heaters among the household appliances that can produce carbon monoxide when badly installed, faulty or poorly maintained[2].
The British Standard that guides gas operatives on using these instruments is BS 7967, a guide for the use of electronic portable combustion gas analysers for measuring carbon monoxide in dwellings and the combustion performance of domestic gas-fired appliances. It covers analysers conforming to BS EN 50379-3, and analysers made to the older BS 7927 remain acceptable under it[3]. Any new analyser you buy should conform to BS EN 50379-3.
Choosing an analyser
| Feature | What to check |
|---|---|
| Standard | Conforms to BS EN 50379-3 for the measurements you need |
| Sensors | Oxygen and carbon monoxide as standard; check whether CO2 is measured or calculated |
| Sensor life and replacement | Expected life, cost and whether you can swap sensors or must send the unit away |
| Probe | Length and fittings to suit the flues and test points you meet, including room CO sampling |
| Records | Bluetooth printing or an app that attaches results to the job record |
| Pressure and temperature | Built-in manometer and flow and return probes save carrying extra tools |
| Calibration service | Turnaround time, loan units and whether there is a local service centre |
| Battery and robustness | Battery life through a long day and resistance to knocks and condensate |
Buy for the appliances you actually work on. A firm that services commercial plant needs different probes and ranges from one that mainly does domestic combis. Ask your merchant or the manufacturer for the specification sheet and check the conformity statement, not just the box.
Calibration: how often and how to schedule it
Analyser sensors and electronics drift over time. Manufacturers set a recalibration interval and it is usually annual. One manufacturer's manual, for example, says that although sensor life is typically more than two years, the analyser should be recalibrated and serviced annually to stop long-term sensor or electronics drift[4]. Follow your own model's instructions, and send it back sooner if it has been dropped, soaked in condensate or gives readings you do not trust.
- Keep each calibration certificate, digitally, against the analyser's serial number.
- Put the due date in the office diary a month ahead, not just on a sticker on the case.
- Stagger due dates across the team so that no more than one analyser is away at a time.
- Book calibration in late spring or summer where you can, not in the first cold week of autumn.
- Ask whether the service centre offers a loan unit or an exchange service, and what it costs.
- Record any repairs or sensor replacements done at calibration.
An analyser out of calibration is not a tool you can use for safety-critical readings. If a certificate has lapsed, take the analyser out of use until it is back, and borrow or hire one rather than carrying on.
Start-up checks every time
Check the certificate date
A quick look at the label and the app before the first job of the day.
Inspect the probe and hose
Look for cracks, kinks and loose connections that would let in air and dilute the sample.
Empty the water trap and check filters
Condensate in the line damages sensors and skews readings. Replace dirty particle filters.
Zero in fresh air
Switch on and let the analyser purge and zero in clean outdoor air, away from flues, vans and engines.
Check the battery
Charge or replace batteries before a long day; low batteries cause errors mid-test.
Confirm the printer or app link
So readings attach to the job rather than being written on a scrap of paper.
Looking after it between calibrations
Most analyser failures are caused by how the instrument is stored and carried rather than by the sensors wearing out. A few habits extend its life and keep readings reliable.
- Store it in its case in the cab, not loose in the back of a van that freezes overnight or bakes in summer.
- Let it purge in fresh air after each test so flue gases do not sit on the sensors.
- Keep the probe and hose dry; hang the hose so condensate drains rather than coiling it wet into the case.
- Carry spare filters and water trap parts, and change them before they look dirty.
- Do not lend analysers between engineers without logging it, so the record of who used which unit stays accurate.
- Report a drop or a soaking straight away, even if the analyser still switches on.
Give one person in the firm the job of owning analysers: certificates, due dates, spares and the loan or hire arrangement. In a small firm that is often the office manager; in a larger one, the technical lead.
Using it on the job
Follow the appliance manufacturer's instructions for where and when to sample, and the analyser manufacturer's instructions for the test sequence. Allow the appliance to run until it is stable before you take a combustion reading, and do not take readings from a cold appliance on a quick test. Where you are investigating a suspected CO problem or a room-sealed appliance with a possible leak, the room CO procedures in the BS 7967 guidance apply; know them and have the probe set-up ready.
If readings are outside the limits the appliance manufacturer sets, follow the industry unsafe situations procedure your training covers. Do not adjust gas rates or combustion settings unless the manufacturer's instructions permit it and you are competent to do so.
Records that stand up later
If a customer, landlord or investigator ever asks what you found, the reading on its own is not enough. They will want to know which analyser was used and whether it was in calibration that day. Make every record traceable.
- Analyser make, model and serial number on every job sheet or certificate.
- Date and time of each reading, and the appliance it relates to.
- The readings themselves, attached digitally rather than retyped.
- Calibration certificate stored centrally with its expiry date.
- Any room CO checks and the result, even if nothing was found.
- Any warning notice issued and what the customer was told.
Carbon monoxide and the customer conversation
Every visit is a chance to check for a working CO alarm. 62% of dwellings in England had a carbon monoxide alarm in 2024, up from 60% in 2023[5], which leaves a large minority without one. Wherever alarms are fitted in the UK they must meet EN 50291-1:2018, although the rules on where alarms are required differ between England, Wales, Scotland and Northern Ireland[6].
Source: Ministry of Housing, Communities and Local Government (GOV.UK), 2024[5]
Show the data
| Item | Value |
|---|---|
| 2023 | 60% |
| 2024 | 62% |
Script: “Your boiler's fine today. I noticed there's no carbon monoxide alarm in here. They're cheap and they warn you if anything changes between services. Would you like me to fit one, or I can tell you what to look for?”
If a customer reports symptoms such as headaches that ease when they leave the house, the NHS advice is to stop using the suspect appliance and get medical advice[2]. Give that advice plainly, and follow your emergency procedures for the appliance.
Worked example: four engineers, four calibration dates
Example: a fictional firm with four engineers used to send all its analysers for calibration in October, the month they were all bought. Each year it hired two analysers while the others were away, at the start of the busiest season. The owner moved the dates to April, May, June and July by sending one analyser early each month in the first year. Calibration certificates now live in the job system, and the office gets a reminder a month before each due date. No engineer is without an analyser in winter, and the firm has stopped hiring.
Mistakes to avoid
- Zeroing the analyser inside the property or next to the van exhaust.
- Ignoring a full water trap until the sensor fails.
- Writing readings by hand and losing the printout.
- Buying the cheapest unit without checking calibration turnaround and loan options.
- Letting one engineer's analyser go out of calibration because the sticker was unreadable.
Related guides: gas smell call handling procedure, thermal imaging cameras and remote boiler diagnostics.
Frequently asked questions
How often should a flue gas analyser be calibrated?
Follow your manufacturer's interval, which is typically annual, and recalibrate sooner if it has been damaged or gives doubtful readings. Keep every certificate.
What standard should my analyser meet?
BS 7967, the guidance for using portable combustion analysers in homes, refers to analysers conforming to BS EN 50379-3. Check the conformity statement for your model.
Can I use an analyser with an expired calibration certificate?
You should not rely on it for safety-critical readings. Take it out of use and borrow or hire a calibrated analyser until yours is back.
Where should I zero my analyser?
In clean outdoor air, away from flues, vehicle exhausts and other combustion sources, following the manufacturer's start-up procedure.
References
legislation.gov.uk (The National Archives) (1998). The Gas Safety (Installation and Use) Regulations 1998, regulation 3 (opens in a new tab) ↑
NHS (2026). Carbon monoxide poisoning (opens in a new tab) ↑
British Standards Institution (catalogue listing, SFS) (2015). BS 7967:2015 Guide for the use of electronic portable combustion gas analysers for the measurement of carbon monoxide in dwellings and the combustion performance of domestic gas-fired appliances (opens in a new tab) ↑
Kane International (hosted by RS Components) (2006). KANE425 flue gas analyser user manual: annual recalibration and service (opens in a new tab) ↑
Ministry of Housing, Communities and Local Government (GOV.UK) (2024). English Housing Survey 2024 to 2025: Chapter 1, housing quality (opens in a new tab) ↑
Heating and Hotwater Industry Council (HHIC) (2025). A guide to carbon monoxide (CO) alarm legislation for heating engineers (opens in a new tab) ↑